Successful off-pump surgical release for extrinsic compression of outflow graft obstruction of a left ventricular assist device

研究成果: Contribution to journalArticle査読

1 被引用数 (Scopus)

抄録

A 36-year-old woman had received support with the Jarvik 2000 left ventricular assist device for 3 years. Cardiac computed tomography revealed outflow graft obstruction, which appeared as external compression of the outflow graft. The patient underwent surgical release of the obstruction after resternotomy. Yellow sediment had become encrusted in the enclosed space between the covering graft and outflow graft and appeared to have been derived from plasma contents. Successful release of the obstruction was achieved by cutting open the covering graft without cardiopulmonary bypass. Neither pump exchange nor outflow graft exchange was required. This off-pump surgery can be a beneficial procedure in terms of less invasiveness and good curability.

本文言語英語
ページ(範囲)402-403
ページ数2
ジャーナルEuropean Journal of Cardio-thoracic Surgery
57
2
DOI
出版ステータス出版済み - 2 1 2020

All Science Journal Classification (ASJC) codes

  • Surgery
  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine

フィンガープリント 「Successful off-pump surgical release for extrinsic compression of outflow graft obstruction of a left ventricular assist device」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル