The Design and Characterization of Dielectric Microcalorimeters for X-ray Photon Detection

T. Kikuchi, M. Hoshino, T. Nakayama, N. Y. Yamasaki, K. Maehata, K. Mitsuda

研究成果: Contribution to journalArticle査読

抄録

The dielectric micro calorimeter (DMC) is a novel radiation detector utilizing a GHz resonator with dielectric thermometer (Sekiya et al. in J Low Temp Phys 167:435, 2012). The advantage of using a DMC is that the detection mechanism is based on a phonon mediation without Johnson noise and quasi-particle decay process. A large format array of DMCs can be easily multiplexed by a resonator circuit in the readout at GHz band width. We describe the design of a DMC as an X-ray photon counter. It is optimized to detect photon at 5.9 keV energy. We consider 18O-doped SrTiO3 (STO18) and Nb-doped KTa( 1 - x )NbxO3 (KTN) as a candidate of dielectric thermometer. Dielectric materials which have sensitivity α(= dlog Cel/ dlog T) are also ideal for our application. We check that both STO18 and KTN have α∼ 10 3 at 100 mK. If we assume a DMC resonator operating at 100 mK, we need a Q value of a resonator to be 2000 for X-ray detection.

本文言語英語
ページ(範囲)250-256
ページ数7
ジャーナルJournal of Low Temperature Physics
184
1-2
DOI
出版ステータス出版済み - 7 1 2016

All Science Journal Classification (ASJC) codes

  • 原子分子物理学および光学
  • 材料科学(全般)
  • 凝縮系物理学

フィンガープリント

「The Design and Characterization of Dielectric Microcalorimeters for X-ray Photon Detection」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。