The first 36 channel fluxgate-sensor-array for the MCG measurement

Hikaru Karo, K. Shimoda, Yoshiaki Maeda, Ichiro Sasada

研究成果: ジャーナルへの寄稿学術誌査読

4 被引用数 (Scopus)

抄録

Fundamental mode orthogonal fluxgate(FM-OFG) is low noise magnetometer because the Barkhausen magnetic noise arising from the core is suppressed by the dc bias current in the excitation. We have established a technique to operate 36 channel FM-OFG sensor array by adding the phase control section to the driving electronics in order to make easy the phase angle adjustment of each of the channel. We have succeed to measure the MCG by using the 36 channel sensor array and a lightweight magnetic shield under the environment of urban magnetic noise. The obtained MCG clearly shows amplitude distributions at the QRS-wave and T-wave similar to those measured by the SQUID MCG measurement system.

本文言語英語
ページ(範囲)224-228
ページ数5
ジャーナルIEEJ Transactions on Sensors and Micromachines
136
6
DOI
出版ステータス出版済み - 1月 1 2016