Two-dimensional magnetic nanoparticle imaging using multiple magnetic sensors based on amplitude modulation

研究成果: Contribution to journalArticle

抜粋

We propose an imaging system for magnetic nanoparticles (MNPs) using multiple magnetic sensors based on amplitude modulation. The system was used to obtain two-dimensional images of two MNP samples spaced 15 mm apart at a depth of 30 mm. An AC magnetic field was used at an operating frequency of 5.4 kHz, and a third-harmonic signal generated by the MNP samples was detected using 16 pickup coils. To increase the position information of the MNP samples, the amplitude of the AC magnetic field was changed from 0.24 to 4.8 mT. The third-harmonic signals were subsequently analyzed using nonnegative least squares to image the MNP samples. The results show that the positions of the two MNP samples were estimated with good accuracy.

元の言語英語
記事番号166765
ジャーナルJournal of Magnetism and Magnetic Materials
505
DOI
出版物ステータス出版済み - 7 1 2020

All Science Journal Classification (ASJC) codes

  • Electronic, Optical and Magnetic Materials
  • Condensed Matter Physics

フィンガープリント Two-dimensional magnetic nanoparticle imaging using multiple magnetic sensors based on amplitude modulation' の研究トピックを掘り下げます。これらはともに一意のフィンガープリントを構成します。

  • これを引用