抄録
The upgraded spectrometer used for charge exchange recombination spectroscopy on the Madison Symmetric Torus resolves emission fluctuations up to 400 kHz. The transimpedance amplifier's cutoff frequency was increased based upon simulations comparing the change in the measured photon counts for time-dynamic signals. We modeled each signal-processing stage of the diagnostic and scanned the filtering frequency to quantify the uncertainty in the photon counting rate. This modeling showed that uncertainties can be calculated based on assuming each amplification stage is a Poisson process and by calibrating the photon counting rate with a DC light source to address additional variation.
本文言語 | 英語 |
---|---|
論文番号 | 11E530 |
ジャーナル | Review of Scientific Instruments |
巻 | 87 |
号 | 11 |
DOI | |
出版ステータス | 出版済み - 11月 1 2016 |
外部発表 | はい |